Ensure switch on the back of the FTIR is ON (not the switch on top)
-
Ensure BioRad USB is plugged into the back of the computer
-
If FTIR is not on, turn it on and wait 15 minutes for the machine to warm up
-
If BioRad USB is missing contact Dr. Harding
-
-
-
Turn on the computer monitor and computer tower
-
-
-
Flip the switch on the front of the FTIR to ON
-
The green LED next to the "resume" and "bot" text should light up. The "laser" LED should light up about 15 seconds after turning on
-
-
-
Open the Spectrum Management program by double clicking the icon on the desktop
-
If FTIR has been idle for a while the program may not open. Fix this by turning the FTIR switch on front OFF then back ON
-
-
-
Double click the Spectra Measurement icon on the left column
-
-
-
Carefully open the lid of the FTIR and twist the black knob counterclockwise and rotate it away from the sample surface
-
Squirt Isopropyl Alcohol onto several kimwipes and lightly wipe the sample surface and the positioning arm tip
-
Close the lid of the FTIR (sample should not be in at this point)
-
-
-
Locate the B icon with the red arrow and click on it
-
Background scan will be run that takes about 2 mins and will produce a graph
-
-
-
Open the blue lid again and lightly place your sample onto the sample surface
-
Rotate the positioning arm clockwise until it is over the sample
-
Twist the black knob clock wise until it touches the sample surface ( you will feel a gentle click)
-
Close the lid of the FTIR
-
-
-
Click the "S" icon with the red arrow through it on the tool bar
-
Wait for 30 seconds for the scan to be taken, a new graph will replace the old graph
-
-
-
A flashing icon depicting a light blue graph will appear on the task bar, click on this icon to open the Spectrum Analysis program
-
Click on the wizard hat icon on the top tool bar on the right and the IR scan will be displayed in the "Know-It-All" software
-
-
-
Move over from the "IR Spectrum" tab to the "Databases" tab within the Search Parameters window
-
Under the "Databases" tab, click on the + icon next to the "Reference-Licensed" option and select "all"
-
On the left middle of the of the screen click on the "Add All" button and the four material categories will be moved to the "Selected for Searching" window
-
Click the "Search" button at the bottom right of the screen
-
Your graph will appear in black in the "MineIt" tool of the software while the orange graph will be the IR scan of the reference material
-
-
-
Select different reference materials in the bottom window to overlay onto your sample spectra
-
The similarity between your sample and the reference sample is given a Hit Quality Index (HQI) score
-
The higher the HQI score, the greater the similarity between the two spectra with the maximum score being 999
-
-
-
The Y-axis can be changed from % Absorbance to % Transmittance by clicking the view tab
-
The spectra can be overlaid, offset, or stacked by clicking on the icons on the toolbar at the top of the "MineIt" window
-
-
-
Take a screenshot of the sample by clicking on the "Snipping Tool" on the bottom toolbar of windows (scissor icon)
-
Save the screenshot onto a USB drive and exit out of the snipping tool window
-
-
-
Open the lid of the FTIR and remove the sample from the FTIR by reversing the steps of placing it in
-
Wipe down the sample surface and the bottom tip of the positioning arm again with isopropyl alcohol on the Kimwipes
-
-
-
Exit out of all the software programs
-
Turn the power switch on the top of the FTIR to "OFF"
-
Ensure that you exit the software before turning off the FTIR
-