Introduction
There are many safety precautions specific to the Bruker Tracer III-SD XRF-Gun. Below is a series of useful steps to follow to ensure the highest degree of safety when handling the Bruker Tracer III-SD XRF-Gun.
-
-
Be sure to follow "ALARA" principles.
-
Take action so X-ray exposure is "As Low As Reasonably Achievable" (ALARA)
-
There is radiation exposure at all 360 degrees of incidence from the analyzer (Picture 2 displays this), albeit less concentrated in some areas.
-
Factors that should be limited are radiation exposure time and distance from radiation. Also shielding from radiation should be maximized.
-
-
-
Always wear a dosimeter when using the XRF gun in order to document radiation absorption.
-
-
-
High Intensity X-rays. Do not expose any part of the body to the beam.
-
X-ray concentration is at its strongest at a 53° incident angle from the examination window to the users' left (as viewed from the users' perspective).
-
This instrument conforms to ANSI N43.3 (Safety Requirements for cabinet X-ray systems)
-
Take special precautions to minimize any radiation exposure if you have ovaries.
-
-
-
X-rays are only generated when the trigger is pulled.
-
A red light illuminates when x-rays are generated.
-
-
-
DO NOT use hands to hold the sample.
-
DO NOT point the analyzer at people during measurement when the red X-ray light is on.
-
-
-